IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 52, 期 12, 页码 2100-2107出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2005.857637
关键词
body movements; health monitoring; heartbeat; noninvasive measurement; pressure sensor; respiration; snoring; vital sign
We have developed a noninvasive pneumatics-based system by which to measure heartbeat, respiration, snoring, and body movements of a subject in bed. A thin, air-sealed cushion is placed under the bed mattress of the subject and the small movements attributable to human automatic vital functions are measured as changes in pressure using a pressure sensor having an almost flat frequency response from 0.1 to 5 kHz and a sensitivity of 56 mV/Pa. Using the newly developed system, heartbeat, respiration, apnea, snoring and body movements are clearly measured. In addition, the optimal signal-to-noise (S/N) ratio by which to evaluate the reliability of the heart rate measurement is presented. Heart rates were measured for four different body postures, 13 different subjects, four different bed mattresses, and three different sensor positions. For these measurements, the S/N ratios ranged from 15.9 to 23.5 dB, and so were determined to be reliable.