4.7 Article Proceedings Paper

Determination of aluminum by electrothermal atomic absorption spectroscopy in lubricating oils emulsified in a sequential injection analysis system

期刊

TALANTA
卷 68, 期 2, 页码 179-186

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2005.06.014

关键词

sequential injection analysis; lubricating oils; emulsions; electrothermal atomic absorption spectrometry; aluminum; on-line analysis

向作者/读者索取更多资源

The sequential injection (SIA) technique was applied for the on-line preparation of an oil in water microemulsion and for the determination of aluminum in new and used lubricating oils by electrothermal atomic absorption spectrometry (ET AAS) with Zeeman-effect background correction. Respectively, 1.0, 0.5 and 1.0 ml of surfactants mixture, sample and co-surfactant (sec-butanol) solutions were sequentially aspirated to a holding coil. The sonication and repetitive change of the flowing direction improved the stability of the different emulsion types (oil in water, water in oil and microemulsion). The emulsified zone was pumped to fill the sampling arm of the spectrometer with a sub-sample of 200 mu l. Then, 10 mu l of this sample solution were introduced by means of air displacement in the graphite tube atomizer. This sequence was timed to synchronize with the previous introduction of 15 jig of Mg(NO3)2 (in a 10 mu l) by the spectrometer autosampler. The entire SIA system was controlled by a computer, independent of the spectrometer. The furnace program was carried out by employing a heating cycle in four steps: drying (two steps at 110 and 130 degrees C), pyrolisis (at 1500 degrees C), atomization (at 2400 degrees C) and cleaning (at 2400 degrees C). The calibration graph was linear from 7.7 to 120 p,g AII(-1). The characteristic mass (mo) was 33.2 pg/0.0044 s and the detection limit was 2.3 mu g AII(-1). The relative standard (RSD) of the method, evaluated by replicate analyses of different lubricating oil samples varied in all cases between 1.5 and 1.7%, and the recovery values found in the analysis of spiked samples ranged from 97.2 to 100.4%. The agreement between the observed and reference values obtained from two NIST Standard Certified Materials was good. The method was simple and satisfactory for determining aluminum in new and used lubricating oils. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据