4.6 Article

Superior Chest Drainage With an Active Tube Clearance System: Evaluation of a Downsized Chest Tube

期刊

ANNALS OF THORACIC SURGERY
卷 91, 期 2, 页码 580-583

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.athoracsur.2010.10.018

关键词

-

资金

  1. United States Army Telemedicine and Advanced Technologies Research Center [W81XWH-05-1-0564]
  2. Global Cardiovascular Innovation Center (Cleveland, OH)

向作者/读者索取更多资源

Purpose. We developed a small-diameter (20-French [Fr]) chest drainage system that incorporates an active tube clearance (ATC) system, and we evaluated its efficacy in an acute hemothorax model. Description. The ATC system includes an internal guidewire that can be advanced into the lumen of the chest tube to keep the tube from clogging. In six pigs, a 20-Fr ATC tube was placed on one side and a 32-Fr standard tube on the other, and 120 mL of fresh blood was infused into each pleural space every 15 minutes for a total of 840 mL. Evaluation. The amount of drainage for 2 hours with a 20-Fr ATC tube was significantly greater than that with the 32-Fr standard tube (525 +/- 179 mL vs 183 +/- 85 mL; p = 0.0032). Conclusions. This is the first time a smaller diameter tube has been shown to have better drainage in the setting of acute bleeding compared with larger diameter tubes. (Ann Thorac Surg 2011;91:580-3) (C) 2011 by The Society of Thoracic Surgeons

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据