Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure

期刊

ANNALS OF THORACIC MEDICINE
卷 6, 期 4, 页码 217-220

出版社

SAUDI THORACIC SOC
DOI: 10.4103/1817-1737.84776

关键词

Non-invasive measurement; PaCO(2); PtCO(2); respiratory failure; SaO(2); SpO(2); transcutaneous electrode

向作者/读者索取更多资源

BACKGROUND: Non-invasive measurement of oxygenation is a routine procedure in clinical practice, but transcutaneous monitoring of PCO(2) (PtCO(2)) is used much less than expected. METHODS: The aim of our study was to analyze the value of a commercially available combined SpO(2)/PtCO(2) monitor (TOSCA-Linde Medical System, Basel, Switzerland) in adult non-invasive ventilated patients with acute respiratory failure. Eighty critically ill adult patients, requiring arterial blood sample gas analyses, underwent SpO(2) and PtCO(2) measurements (10 min after the probe was attached to an earlobe) simultaneously with arterial blood sampling. The level of agreement between PaCO(2) - PtCO(2) and SaO(2) - SpO(2) was assessed by Bland-Altman analyses. RESULTS: Both, SaO(2) from blood gas analysis and SpO(2) from the transcutaneous monitor, and PaCO(2) and PtCO(2) were equally useful. No measurements were outside of the acceptable clinical range of agreement of +/- 7.5 mmHg. CONCLUSIONS: The accuracy of estimation of the TOSCA transcutaneous electrode (compared with the gold standard blood sample gas analysis) was generally good. Moreover, TOSCA presents the advantage of the possibility of continuous non-invasive measurement. The level of agreement of the two methods of measurement allows us to state that the TOSCA sensor is useful in routine monitoring of adults admitted to an intermediate respiratory unit and undergoing non-invasive ventilation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据