An accuracy evaluation of the T-Line® Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients

期刊

ANESTHESIA AND ANALGESIA
卷 102, 期 2, 页码 484-490

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1213/01.ane.0000194873.52453.bd

关键词

-

向作者/读者索取更多资源

Continuous beat-to-beat arterial blood pressure (BP) monitoring with a simultaneous arterial waveform display is typically achieved with an invasive arterial catheter. We evaluated a noninvasive device, the T-Line (R) Tensymeter, that provides a calibrated arterial pressure waveform from which continuous BP measurements and heart rate may be computed by either a bedside host monitor or the tensymeter device itself. In 25 patients given general anesthesia, we measured systolic, mean, and diastolic BPs via the tensymeter and compared these measurements with those obtained from the contralateral radial artery catheter. Data were analyzed using the Bland Altman test to determine agreement between the two systems. The mean +/- SD bias and precision (mm Hg) were as follows: 1.7 +/- 7.0 and 5.7 +/- 4.4 for systolic BP; 2.3 +/- 6.9 and 5.7 +/- 4.5 for diastolic BP; and 1.7 +/- 5.3 and 4.0 +/- 4.8 for mean BP. Noninvasive pressures from the tensymeter-produced arterial waveform agreed with simultaneous contralateral BPs measured from arterial catheters within an acceptable clinical range for a limited population of surgical patients studied over a systolic arterial BP range from 41 to 189 mm Hg without significant temporal performance degradation. The tensymeter may enable physicians to circumvent arterial cannulation in certain circumstances (such as with low- or intermediate-risk procedures) on patients when beat-to-beat BP measurement is desirable.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据