4.6 Article

Air ionisation and colonisation/infection with methicillin-resistant Staphylococcus aureus and Acinetobacter species in an intensive care unit

期刊

INTENSIVE CARE MEDICINE
卷 32, 期 2, 页码 315-317

出版社

SPRINGER
DOI: 10.1007/s00134-005-0002-8

关键词

acinetobacter; healthcare-associated infection; ionisation; nosocomial infection; Staphylococcus aureus

向作者/读者索取更多资源

Objective: To determine effect of negative air ions on colonisation/infection with methicillin-resistant Staphylococcus aureus (MRSA) and Acinetobacter species in an intensive care unit. Design: Prospective single-centre cross-over study in an adult general intensive care unit. Patients: 201 patients whose stay on the unit exceeded 48 hour's duration. Intervention:Six negative air ionisers were installed on the unit but not operational for the first 5 months of the study (control period). Devices were then operational for the following 5.5 months. Measurements and results:30 and 13 patients were colonised/infected with MRSA and Acinetobacter spp., respectively, over 10.5 months. No change in MRSA colonisation/infection was observed compared with the 5 month control period. Acinetobacter cases were reduced from 11 to 2 (p = 0.007). Conclusion:Ionisers may have a role in the prevention of Acinetobacter infections.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据