4.7 Article

Supported capillary membrane sampling-gas chromatography on a valve with a pulsed discharge photoionization detector

期刊

ANALYTICA CHIMICA ACTA
卷 560, 期 1-2, 页码 197-206

出版社

ELSEVIER
DOI: 10.1016/j.aca.2005.12.015

关键词

pulsed discharge photoionization detector; gas chromatography; water analysis; trihalomethanes (THM4); disinfection by-products

向作者/读者索取更多资源

The objective of this study was to develop and evaluate a simple miniaturized approach to gas chromatography termed supported capillary membrane sampling-gas chromatography on a valve (SCMS-GCOV). The prototype instrument uses a pulsed discharge photoionization detector (PDPID) as the GC detector. Two different analyzers were constructed. The first used a bench scale SCMS-GC. The second was a miniaturized SCMS-GCOV analyzer. The SCMS-GCOV analyzer was used to monitor the concentrations of trihalomethanes (THM4) in drinking water. Details are given for the construction of the SCMS-GC-PDPID and the SCMS-GCOV-PDPID analyzers. The results of method detection limit, accuracy and precision studies are reported for analysis of THM4 and are compared to other methods of THM4 analysis. Side-by-side comparison studies are conducted between the SCMS-GCOV and USEPA 502.2 with promising results. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据