期刊
ARTIFICIAL CELLS BLOOD SUBSTITUTES AND BIOTECHNOLOGY
卷 34, 期 2, 页码 223-239出版社
INFORMA HEALTHCARE
DOI: 10.1080/10731190600581775
关键词
creatine; enzyme electrode; amperometry; poly(vinylferrocen); creatinase; sarcosine oxidase
A new enzyme electrode for the determination of creatine was developed by immobilizing creatinase (CI) and sarcosine oxidase (SO). The enzymes were co-immobilized in a poly(vinylferrocenium) matrix onto the surface of a platinum working electrode. Crosslinking with glutaraldehyte (GA) and bovine serum albumin (BSA) was selected as the best immobilization method for the enzymatic system. Determination of creatine was performed by the oxidation of enzymatically generated H2O2 at +0.7 V vs. Ag/AgCl. The linear working range of the electrode was 2.0 x 10(-5) -3.2 x 10(-4) M and the response time was about 50 s. The effects of pH, temperature, enzyme ratio and buffer concentration were investigated and optimum parameters were found to be 7.5, 37 degrees C, 2.5:1 (CI:SO) and 0.05 M, respectively. The stability and reproducibility of the enzyme electrode have been also studied.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐