Fast low-angle dual spin-echo (FLADE): A new robust pulse sequence for structural imaging of trablecular bone

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 55, 期 3, 页码 465-471

出版社

WILEY
DOI: 10.1002/mrm.20789

关键词

FLADE; FLASE; trabecular bone; double spin echo; high resolution imaging

资金

  1. NIAMS NIH HHS [R01 AR41443, R01 AR49553] Funding Source: Medline
  2. NIBIB NIH HHS [9T32EB000814] Funding Source: Medline

向作者/读者索取更多资源

Mechanical strength and fracture resistance of trabecular bone (TB) are largely determined by the structural arrangement of individual trabeculae. Fast 3D spin-echo approaches are preferable to gradient echoes in that they are less sensitive to local induced gradients at the bone/marrow interface caused by magnetic susceptibility difference between the two tissues. FLASE is a 3D pulse sequence that serves this purpose. Here, we present a new pulse sequence dubbed FLADE (fast low-angle dual spin-echo) that overcomes some of the limitations inherent to FLASE, such as sensitivity to artifactual stimulated echoes. The double-echo sequence features a flip angle < 90 degrees allowing for TR << T-1. The second phase-reversal pulse has the dual function of creating a second echo and restoring inverted longitudinal magnetization. The prolonged TR, made possible by sampling only half of k(z)-space, is used to collect navigator echoes in adjacent slabs for sensing subpixel translational displacements. FLADE is shown to provide SNR comparable to FLASE while having narrower point-spread function and being more robust to imperfections in the nonselective 180 degree pulses. Structural parameters derived from the in vivo images with the two pulse sequences are highly correlated, therefore suggesting that clinical data obtained with either pulse sequence can be merged.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据