期刊
PROTEOMICS
卷 6, 期 5, 页码 1427-1436出版社
WILEY
DOI: 10.1002/pmic.200500078
关键词
cancer; PAI-1; protein microarrays; uPA; VEGF
A multiplexed fluorescence immunoassay using a novel planar waveguide technology-based microarray system, ZeptoMARK (Zeptosens), was developed to detect simultaneously urokinasetype plasminogen activator (uPA), plasminogen activator inhibitor-1 (PAI-1), and vascular endothelial growth factor (VEGF) in extracts of breast cancer tissues. The three analytes assay was cross-validated with single-analyte ELISA/chemiluminescence immunosorbent assay tests, revealing good correlations and enhanced assay sensitivities (LODs) of 1 pg/mL for uPA, 33 pg/ mL for PAI-1, and 1 pg/mL for VEGF. Values were well within the 80-120% limits for assay recovery and within the +/- 20% limits for assay precision. The uPA, PAI-1, and VEGF results obtained from 50 breast cancer cytosols using the protein array system demonstrated that the microarray-based multiplexed assay is a sensitive and robust tool to be used for the simultaneous quantification of cancer markers in small breast cancer tissue samples (core biopsies). The miniaturized, multiplexed assay format has a potential to be used for the quantitative analysis of a larger set of validated markers with significance in disease management.