4.8 Article

Micromachined sensor for lactate monitoring in saliva

期刊

BIOSENSORS & BIOELECTRONICS
卷 21, 期 9, 页码 1770-1776

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2005.09.015

关键词

lactate sensor; saliva; amperometric measurement; silicon enzyme sensors; iridium reference electrode

向作者/读者索取更多资源

A miniaturised sensor for continuous lactate measurement in saliva was developed and tested. The sensor was built using silicon microfabrication technologies. The size of the chip is 5.5 rnin x 6.4 min x 0.7 mm and features a working, a counter and an Iridium reference electrode. The chip has a cavity whose floor is perforated by fine pores. The cavity contains the enzyme lactate oxidase (LOD), which is immobilised in an agarose gel. Prior to the amperometric detection of the reaction product hydrogen peroxide at the working electrode, the analyte lactate has to pass the pores to reach the cavity with the lactate oxidase by diffusion. To test the silicon sensor, capillary blood and saliva samples were obtained during standardised ergometer tests. Salivary lactate concentrations were determined with the sensor and compared to photo metrically derived data from a lab-automate. In addition the saliva data were compared to standard capillary blood lactate concentrations measured with a pocket photometer. Lactate concentration versus load graphs were plotted and compared visually showing very similar progressions. The novel approach enables a location independent, permanent real-time measurement of the lactate concentration during exercise. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据