4.1 Article

Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 29, 期 4, 页码 380-385

出版社

BLACKWELL PUBLISHING
DOI: 10.1111/j.1540-8159.2006.00357.x

关键词

electromagnetic interference; home monitoring; pacemaker; mobile phones

向作者/读者索取更多资源

The aim of this study was to evaluate whether global system for mobile communication (GSM) cellular phones can affect the home monitoring (HM) function implemented in last generation pacemakers (PM). Methods: We performed in vitro and in vivo tests on the BA03 PM (Biotronik GmbH, Berlin, Germany). In vitro we evaluated whether an improper HM procedure or an altered patient-activated HM function occurred. We used two cellular phone models, with a fixed external or internal antenna, positioned close to both the PM and the mobile phone-like device, during handover, ringing, and talking. All the tests were done with the PM in air, at 900 and 1,800 MHz GSM bonds, under worst case conditions. A subset of these tests was repeated in 17 patients: the mobile phones were moved both around the PM implant site and the mobile cell phone-like device, during talking. Results: In vitro, neither the HM procedure nor PM functioning were corrupted by the GSM communications: all the transmissions were correctly received, with a maximum transmission delay of about 110 seconds. In vivo, the rate of successful trasmissions was 93%. Conclusion: Our data show that HM function does not call for specific restrictions on the use of GSM cellular phones.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据