Silver solid amalgam electrodes as sensors for chemical carcinogens

期刊

SENSORS
卷 6, 期 4, 页码 445-452

出版社

MOLECULAR DIVERSITY PRESERVATION INTERNATIONAL
DOI: 10.3390/s6040445

关键词

solid amalgam electrodes; voltammetry; carcinogens; 3-nitrofluoranthene; Ostazine Orange

向作者/读者索取更多资源

The applicability of differential pulse voltammetry (DPV) and adsorptive stripping voltammetry (AdSV) at a non-toxic meniscus-modified silver solid amalgam electrode (m-AgSAE) for the determination of trace amounts of genotoxic substances was demonstrated on the determination of micromolar and submicromolar concentrations of 3-nitrofluoranthene using methanol - 0.01 mol L-1 NaOH ( 9: 1) mixture as a base electrolyte and of Ostazine Orange using 0.01 mol L-1 NaOH as a base electrolyte.