Transfer of NIR calibrations for pharmaceutical formulations between different instruments

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2005.10.042

关键词

calibration transfer; local centring; slope and bias correction; piecewise direct standardisation; dispersive instrument; Fourier transform instrument

向作者/读者索取更多资源

In order to evaluate how well existing techniques for transferring NIR calibrations perform for solid pharmaceutical formulations, a study oil four assays of active ingredients was undertaken. The study included two configurations of dispersive NIR instruments and one Fourier transform (FT) instrument. Three methods for calibration transfer: slope/bias correction, local centring and piecewise direct standardisation (PDS), were tested and evaluated. Our conclusions are that the calibration transfer methods tested can perform equally well. It was shown that it is possible to transfer calibrations between instruments of different configurations or even of different types, without loosing the prediction ability of the calibration. To achieve a good calibration transfer, a larger variation in the content of the active ingredient in the samples and more samples are needed for the slope and bias correction method compared to the local centring method. For PDS to be a successful calibration transfer method, an optimisation of the number of transfer samples and how they are selected together with various factors specific for this method is needed. Local centring is the preferred transfer method as its performance is excellent yet it is simple to perform, no optimisation is needed, only a few transfer samples are required and the transfer samples do not have to vary in their content of the active ingredient. (c) 2005 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据