4.0 Article

Imaging Optimization of Temporal Bones With Cochlear Implant Using a High-resolution Cone Beam CT and the Corresponding Effective Dose

期刊

ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY
卷 124, 期 6, 页码 466-473

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/0003489414565004

关键词

cochlear implant; inner ear trauma; radiology; middle ear

资金

  1. EC [281056]

向作者/读者索取更多资源

Objective: To evaluate the impact of tube voltage, tube current, pulse number, and magnification factor on the image quality of a novel experimental set-up and the corresponding radiation. Materials and Methods: Six human temporal bones with cochlear implant were imaged using various tube voltages, tube currents, pulse numbers, and magnification. The effect of radiation was evaluated using a metaloxide semiconductor field-effect transistor (MOSFET) dosimeter device on an anthropomorphic RANDO RAN102 male head phantom. A copper and aluminum combination filter was used for hardware filtration. Results: Overall, 900 frames, 11 mA, and 88 kV provided the best image quality. In temporal bones imaged with the optimized parameters, the cochlea, osseous spiral lamina, modiolus, stapes, round window niche, and oval window landmarks were demonstrated with anatomic structures still fully assessable in all parts and acceptable image quality. The most dominant contributor to the effective dose was bone marrow (36%-37 %) followed by brain (34%-36%), remainder tissues (12%), extra-thoracic airways (7%), and oral mucosa (5%). Conclusions: By increasing the number of frames, the image quality of the inner ear details obtained using the novel cone-beam computed tomography improved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据