4.8 Article

Determination of Os and Re isotope ratios at subpicogram levels using MC-ICPMS with solution nebulization and multiple ion counting

期刊

ANALYTICAL CHEMISTRY
卷 78, 期 11, 页码 3794-3799

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac060183t

关键词

-

向作者/读者索取更多资源

A precise and accurate determination method of Os-187/ Os-188, Os-189/Os-188, and Re-185/Re-187 ratios of down to 0.2 pg of Os and 0.08 pg of Re using multicollector inductively coupled plasma mass spectrometry (MC-ICPMS) with multiple ion counting has been developed. Os and Re were introduced into MC-ICPMS as 0.5 mol L-1 HF solutions through the desolvator, and Re-185, Os-187, Os-188, and Os-189 ions were detected simultaneously by four channeltrons. The Os and Re ratios were determined by the standard bracketing method, in which channeltron yields and mass discrimination factors are corrected together. Os and Re memories after 800-s wash were < 0.1% and negligible, typically 0.02 and 0.03%, respectively. Isotope ratios of Os-187/ Os-188, Os-189/Os-188, and Re-185/Re-187 were constant within error in the concentration ranges of 1-100, 1-40, and 0.4-5 pg mL(-1) with reproducibility (1 sigma) of 2.7-0.14, 0.33-0.10, and 0.41-0.19%, respectively. For analysis of larger amounts of Os and Re, the Faraday cup measurement was employed. The precision and reproducibility obtained in this study are comparable to those of N-TIMS and better than MC-ICPMS achieved so far with a capability of higher sample throughput with simpler sample preparation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据