A precise and accurate determination method of Os-187/ Os-188, Os-189/Os-188, and Re-185/Re-187 ratios of down to 0.2 pg of Os and 0.08 pg of Re using multicollector inductively coupled plasma mass spectrometry (MC-ICPMS) with multiple ion counting has been developed. Os and Re were introduced into MC-ICPMS as 0.5 mol L-1 HF solutions through the desolvator, and Re-185, Os-187, Os-188, and Os-189 ions were detected simultaneously by four channeltrons. The Os and Re ratios were determined by the standard bracketing method, in which channeltron yields and mass discrimination factors are corrected together. Os and Re memories after 800-s wash were < 0.1% and negligible, typically 0.02 and 0.03%, respectively. Isotope ratios of Os-187/ Os-188, Os-189/Os-188, and Re-185/Re-187 were constant within error in the concentration ranges of 1-100, 1-40, and 0.4-5 pg mL(-1) with reproducibility (1 sigma) of 2.7-0.14, 0.33-0.10, and 0.41-0.19%, respectively. For analysis of larger amounts of Os and Re, the Faraday cup measurement was employed. The precision and reproducibility obtained in this study are comparable to those of N-TIMS and better than MC-ICPMS achieved so far with a capability of higher sample throughput with simpler sample preparation.