4.7 Article

Determination of Se(IV) by anodic stripping voltammetry using gold electrodes made from recordable CDs

期刊

TALANTA
卷 69, 期 4, 页码 877-881

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2005.11.031

关键词

selenium; gold electrode; recordable CDs

向作者/读者索取更多资源

This work presents a simple, reproducible and low cost method, employing differential pulse anodic stripping voltammetry, for determination of selenium(IV). A gold electrode obtained from recordable compact disks (CD-R) was used to evaluate the voltammetric behavior of the metallic ion in 0.1 mol L-1 HClO4. To evaluate the voltammetric behavior of Se(IV), parameters such as deposition potential and deposition time were optimized. A wide linear response range, from 0.5 to 291 ng mL(-1), was obtained using a 5.0 mm, diameter gold electrode. Recovery tests for Se(IV) utilizing standard reference solutions provided values between 94 and 96%. (c) 2005 Elsevier B.V All rights reserved.