Objective The objective of this study was to assess differences in the semiquantitative values of F-18-fluorodeoxyglucose (F-18-FDG) uptake among different positron emission tomographic (PET) systems. Methods A phantom study was performed to compare standardized uptake value (SUV) in five PET scanners including a dedicated PET scanner and four PET/computed tomography (CT) scanners. Radioactivity simulating the SUV of 2.5 was filled in the hot spheres (8 mm, 11 mm, 14 mm, 18 mm, 22 mm, and 27 mm) that were set in the cylindrical phantom with the background SUV of 1.0. Data acquisition and reconstruction were performed according to routine and standardized conditions. The standardized condition was as follows: CT acquisition (120 kVp, 50 mA) and PET acquisition (2-min acquisition with a slice thickness of 2 mm); reconstruction was performed by ordered subsets expectation maximization + Fourier rebinning. Detectability of hot spheres and SUV was compared between routine condition and standardized condition with five PET scanners. Results On routine condition, two cameras could detect a 14-mm sphere clearly. On the other hand, the visualization of hot spheres by the standardized condition was remarkably variable. Semiquantitative evaluation revealed that a maximum of 45.7% error was recognized with the 27-mm sphere by the routine condition, although the standardized condition could reduce the error to 22.6%. Conclusions Detectability depends not only on the PET machine but also on the imaging protocol. The results indicate that SUV is variable with PET machines under routine conditions of data acquisition and reconstruction. Standardization of the reconditions can reduce variability and maximum difference in the SUV by half.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据