4.3 Article

Standards, calibration, and guidelines in microcalorimetry - Part 2. Calibration standards for differential scanning calorimetry - (IUPAC Technical Report)

期刊

PURE AND APPLIED CHEMISTRY
卷 78, 期 7, 页码 1455-1476

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1351/pac200678071455

关键词

IUPAC physical and biophysical chemistry division; microcalorimetry; DSC; heat capacity; enthalpy measurements

向作者/读者索取更多资源

Differential scanning calorimeters (DSCs) are widely used for temperature, heat capacity, and enthalpy measurements in the range from subambient to high temperatures. The present recommendations describe procedures and reference materials (RMs) for the calibration of DSCs. The recommendations focus on the calibration of the response of the instrument and on the estimation of the measurement uncertainty. The procedures for temperature, enthalpy, and heat-flow rate calibration are given in detail. Calibration on cooling has also been considered. Recommended RMs are listed, and the relevant properties of these materials are discussed.