4.8 Article

Mechanical and histological evaluation of a PMMA-based bone cement modified with γ-methacryloxypropyltrimethoxysilane and calcium acetate

期刊

BIOMATERIALS
卷 27, 期 21, 页码 3897-3903

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.biomaterials.2006.03.002

关键词

bioactivity; bone cement; mechanical property; in vitro test; in vivo test

向作者/读者索取更多资源

Polymethylmethacrylate (PMMA) bone cement is widely used for prosthetic fixation in orthopaeclic surgery, however, the interface between bone and cement is a weak zone. We developed a bioactive PMMA cement through modification with gamma-methacryloxypropyl-trimethoxysilane (MPS) and calcium acetate. The purpose of this study was to compare the handling, mechanical and histological properties of the modified bone cement with those of the conventional cement. The modified specimens exhibited higher bonding strength between bone and implant. Histological observation and micro-focus X-ray computed tomogram (micro-CT) images showed that the modified cement exhibited osteoconduction, which the conventional PMMA bone cement lacked. The modification was found to be effective in enabling osteoconduction with PMMA bone cement, thus providing stable fixation for a long period after implantation. (c) 2006 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据