4.3 Article

Badge-type diffusive sampler using 3-methyl-2-benzothiazolinone hydrazone for measuring formaldehyde in indoor air

期刊

ANALYTICAL SCIENCES
卷 22, 期 7, 页码 993-997

出版社

JAPAN SOC ANALYTICAL CHEMISTRY
DOI: 10.2116/analsci.22.993

关键词

-

向作者/读者索取更多资源

The evaluation of a badge-type diffusive sampler for measuring formaldehyde using 3-methyl-2-benzothiazolinone hydrazone (MBTH) was investigated. On average, the formaldehyde concentration in blanks was reduced by approximately 31% by cleaning procedures. The cleaning techniques did not significantly differ in effectiveness. The maximum sampling rate was 22.4 +/- 3.5 mL min(-1) at MBTH concentrations of 0.05%. The formaldehyde concentration in blanks did not appreciably increase over a period of about 1 month at room temperature, and was 0.36 +/- 0.03 mu g, with a relative standard deviation of 8%. The diffusive sampler had good precision and accuracy for measuring formaldehyde in indoor environments. For a 24-h exposure time, the limits of detection and quantification calculated with the field blanks were 9.7 and 13.8 ppb, respectively. The minimum exposure times were calculated based on the measured and calculated limits of quantification, the sampling rate, and the atmospheric formaldehyde concentration. The capacity of the diffusive sampler with 0.5% MBTH was 3 ppm h(-1), approximately 1.5-times the capacity when the MBTH concentrations were 0.05%.