4.6 Article Proceedings Paper

Comparison of femoral head penetration using RSA and the Martell method

期刊

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/01.blo.0000224018.88410.83

关键词

-

向作者/读者索取更多资源

Radiostereometry has high precision and accuracy measuring polyethylene wear in total hip arthroplasty but requires a specialized setup. The Martell method is simpler and can be used on larger populations. The hypothesis that the radiostereometry analysis and the Martell analysis would yield comparable wear data from the same group of patients having total hip arthroplasty was tested. A group of twenty-five total hip arthroplasty patients who had both radiostereometry and standard anterior-posterior pelvic and cross-table lateral radiographs of sufficient quality for analysis were identified. The films were taken at postoperative periods of 6 weeks, 1 year, 2 years, and 5 years. Femoral head penetration was measured by both methods at each time point. The median penetration rates measured by each method decreased over time. Penetration results were affected by method of analysis, time, and dimension, with greater penetration for Martell compared with radiostereometry at each time point, greater penetration with increasing time for each method, and larger three-dimensional magnitude compared with two-dimensional analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据