期刊
CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 44, 期 8, 页码 987-990出版社
WALTER DE GRUYTER & CO
DOI: 10.1515/CCLM.2006.178
关键词
homocysteine; HPLC; tandem mass spectrometry (MS/MS); routine practice
Background: Total plasma homocysteine has emerged as an independent risk factor for vascular disease. To meet increasing requests by clinicians for this homocysteine determination, a rapid assay for a routine use has been developed. Methods: A robust, stable, isotope-dilution liquid chromatography tandem mass spectrometry (LC/MS/MS) method is described, including all the practical details and analytical performance results. Results: The method allows homocysteine quantitation over a linear working range up to 100 mmol/L, with the limit of quantification estimated at a low value of 0.09 mmol/L. Total analytical imprecision is less than 4%. Accuracy was assessed by measuring the homocysteine concentration in a serum Standard Reference Material. Conclusions: The method was demonstrated to be quick, reliable and cheap after 1 year of use on a time-shared instrument in our hospital unit.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐