4.7 Article Proceedings Paper

Flow screen-printed amperometric detection of p-nitrophenol in alkaline phosphatase-based assays

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 385, 期 7, 页码 1202-1208

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-006-0367-8

关键词

alkaline phosphatase; p-nitrophenyl phosphate/p-nitrophenol; screen-printed carbon electrode; flow-injection analysis

向作者/读者索取更多资源

p-Nitrophenyl phosphate is one of the most widely used substrates for alkaline phosphatase in ELISAs because its yellow, water-soluble product, p-nitrophenol, absorbs strongly at 405 nm. p-Nitrophenol is also electroactive; an oxidative peak at 0.97 V (vs. an Ag pseudoreference electrode) is obtained when a bare screen-printed carbon electrode is used. When an amperometric detector was coupled to a flow-injection analysis system the detection limit achieved for p-nitrophenol was 2 x 10(-8) mol L-1, almost two orders of magnitude lower than that obtained by measuring the absorbance of the compound. By use of this electrochemical detection method, measurement of 7 x 10(-14) mol L-1 alkaline phosphatase was achieved after incubation for 20 min. The feasibility of coupling immunoassay to screen-printed carbon electrode amperometric detection has been demonstrated by performing an ELISA for detection of pneumolysin, a toxin produced by Streptococcus pneumoniae, which causes respiratory infections. The method is simple, reproducible, and much more sensitive than traditional spectrophotometry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据