This paper presents design and implementation of a wireless pressure sensor system for biomedical application. The system consists of a front-end Micro-Electro- Mechanical System (MEMS) sensing capacitor along with an optimised MEMS-based oscillator for signal conditioning circuit. In this design, vertical fringed comb capacitor is employed due to the advantages of smaller area, higher linearity and larger full scale change in capacitance compared to parallel plate counterparts. The MEMS components are designed in Coventorware design suite and their Verilog-A models are extracted and then imported to Cadence for co-simulation with the CMOS section of the system using AMI 0.6-micron CMOS process. In this paper, an optimisation method to significantly reduce the system power consumption while maintaining the system performance sufficient is also proposed. A phase noise optimisation approach is based on the algorithm to limit the oscillator tail current. Results show that for the pressure range of 0-300 mmHg the device capacitance range of 1.31 pF - 1.98 pF is achieved which results in a frequency sweep of 2.54 GHz - 1.95 GHz. Results also indicate that a 42% reduction of power consumption is achieved when the optimisation algorithm is applied. This characteristic makes the sensor system a better candidate for wireless biomedical applications where power consumption is the major factor.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据