Reliability of air displacement plethysmography in a large, heterogeneous sample

期刊

MEDICINE AND SCIENCE IN SPORTS AND EXERCISE
卷 38, 期 8, 页码 1505-1509

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1249/01.mss.0000228950.60097.01

关键词

Bod Pod (R); densitometry; body composition; body volume

向作者/读者索取更多资源

Purpose: This study was conducted to deter-mine the reliability of ADP using the Bod Pod (R) in a large, heterogeneous sample. Methods: A total of 980 healthy men and women (30 +/- 15 yr, mean +/- SD) completed two body composition assessments separated by 15-30 min. All testing was done in accordance with the manufacturer's instructions. Results: A significant correlation (r = 0.992, P = 0.001) was found between body density (BD) 1 (1.046 +/- 0.001 kg(.)L(-1); mean +/- SEM) and BD 2 (1.046 +/- 0.001 kg(.)L(-1)). A paired t-test revealed no significant difference between BD I and 2 (P = 0.935). The coefficient of variation (CV) for BD was 0.15%. A significant intraclass correlation coefficient (ICC) was found for BD (ICC = 0.996, P 0.001), and the standard error of measurement (SEM) was 0.001 kg(.)L(-1). Body mass (BM) I and 2 were correlated significantly (r = 0.999, P = 0.001); however, a significant (P = 0.001) decrease was seen from BM 1 (75.510 +/- 0.461 kg) to BM 2 (75.497 +/- 0.461 kg). Body volume (BV) tended to decrease (P = 0.08) from BV 1 (69.900 +/- 0.449 L) to BV 2 (69.884 +/- 0.449 L). Conclusion: ADP using the Bod Pod (R) appears to assess BD reliably; however, the observed CV suggests that multiple trials are necessary to detect small treatment effects.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据