期刊
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM
卷 91, 期 8, 页码 3055-3061出版社
ENDOCRINE SOC
DOI: 10.1210/jc.2006-0710
关键词
-
Context: We have recently introduced liquid chromatography-tandem mass spectrometry (LC-MS/MS) for 25-hydroxyvitamin D-2 (25OHD(2)) and 25OHD(3) testing. During subsequent clinical use, we identified significantly elevated results in some infants. We hypothesized this might represent assay interference caused by C-3 epimers of 25OHD2 or 25OHD3. Objective: Our aims were to 1) determine the prevalence of C-3 epimers of 25OHD2 or 25OHD3 in human serum, and 2) identify the patient populations that might be affected. Study Design: We modified our LC-MS/MS method to allow detection of C-3 epimers. We retested specimens from four patient groups with the new method and an extracted RIA: 1) children less than 1 yr old, 2) children 1-18 yr old, 3) adults aged 20-87 yr with liver disease, and 4) adults aged 19-91 yr without liver disease. Results: In 172 children from group 1 with detectable 25OHD2 or 25OHD3, we identified C-3 epimers in 39 (22.7%). The epimers contributed 8.7-61.1% of the total 25-OHD. There was an inverse relationship between patient age and epimer percentage (r = 0.48; P < 0.002). The RIA gave accurate 25-OHD results that correlated with the modified LC-MS/MS method. No C-3 epimers were detected in any of the other groups. Conclusions: Significant concentrations of C-3 epimers of 25OHD2 or 25OHD3 are commonly found in infants. This can lead to overestimation of 25-OHD levels. Measurements in children less than 1 yr should therefore be performed with an assay that allows accurate detection of 25- OHD in the presence of its C-3 epimers.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐