Automated measurement of mean wall thickness in the common carotid artery by MRI: A comparison to intima-media thickness by B-mode ultrasound

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 24, 期 2, 页码 379-387

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/jmri.20636

关键词

carotid artery MRI; atherosclerosis; intima-media thickness; stastical shape modelling; B-mode ultrasound

资金

  1. NHLBI NIH HHS [T32 HL07828] Funding Source: Medline

向作者/读者索取更多资源

Purpose: To determine whether the mean wall thickness (MWT) of the common carotid artery (CCA) measured by MRI is comparable to B-mode ultrasound (US) measurement of the intima-media thickness (IMT), an established marker of cardiovascular risk. Materials and Methods: As part of the two-year ORION trial, 43 patients with 16-79% stenosis by duplex US underwent high-resolution MRI and B-mode US examinations of their carotid arteries. Twenty-eight carotid arteries were identified as having both sufficient proximal coverage and adequate image quality of the CCA on MRI and a corresponding US. A novel algorithm utilizing statistical shape modeling was developed to automatically detect and measure MWT to within subpixel accuracy. The interrater and interscan reproducibility of the MWT measurement was computed as the root-mean-square (RMS) difference. The MWT and IMT measurements were compared via the Pearson correlation coefficient. Results: The MWT and IMT had a high Pearson correlation coefficient (r = 0.93; P < 0.001). The RMS difference between readers and between scans was 0.01 mm and 0.04 mm,respectively. Our automated algorithm correctly identified the lumen in 28 cases (100%) and the outer-wall boundary in 26 cases (93%). Conclusion: Automated measurements of the MWT by MRI are reproducible and have a high correlation with the IMT by B-mode US.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据