4.6 Article

A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630nm

期刊

OPTICS EXPRESS
卷 14, 期 16, 页码 7420-7435

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.14.007420

关键词

-

类别

向作者/读者索取更多资源

The lack of a primary method for determination of optical parameters remains a significant barrier in optical study of turbid media. We present a complete system of experimental setups and Monte Carlo modeling tools for fast and accurate solution of the inverse problem from the measured signals of homogeneous turbid samples. The calibration of the instrument and validation of the Monte Carlo modeling have been carried out to ensure the accuracy of the inverse solution. We applied this method to determine the optical parameters of turbid media of 10% intralipid between 550 and 940nm and 20% intralipid between 550 and 1630nm. (c) 2006 Optical Society of America.