Comparison of ELISA and capillary electrophoresis with laser-induced fluorescence detection in the analysis of Ochratoxin A in low volumes of human blood serum

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2006.04.048

关键词

capillary electrophoresis; CE-LIF human blood serum; low sample volume; Ochratoxin A; ELISA

向作者/读者索取更多资源

In this paper the determination of Ochratoxin A (OTA) in low volumes of human blood serum by enzyme-linked immunosorbent assay (ELISA) is compared with an appropriate capillary electrophoresis with laser-induced fluorescence detection (CE-LIF) method. In order to use ELISA for high-throughput analysis in epidemiological studies no sample cleanup was performed. Both methods showed a limit of detection (LOD) of 0.5 ng/mL. Comparing the precisions of both methods, the data show that the quantified concentrations in ELISA are higher than the corresponding concentrations in the CE-LIF method. Using a matrix calibration curve instead of a standard calibration curve the reproducibilities of both methods are comparable. No additional matrix effect could be observed by adding phenylalanine as probable matrix compound to the serum. (c) 2006 Elsevier B.V. All rights reserved.