4.7 Article

Determination of glutathione in hemolysed erythrocyte with amperometric sensor based on TTF-TCNQ

期刊

CLINICA CHIMICA ACTA
卷 371, 期 1-2, 页码 152-158

出版社

ELSEVIER
DOI: 10.1016/j.cca.2006.03.006

关键词

glutathione determination; hemolysed erythrocyte; TTF-TCNQ

向作者/读者索取更多资源

Background: GSH has a relevant role in human metabolism as an indicator of disease risks. An amperometric sensor for glutathione (GSH) determination is described as an alternative method featuring simple construction procedure and short time analysis. Method: The developed sensor was used to determine glutathione at low potential using a TTF-TCNQ complex. Results: The sensor exhibits a linear response range from 5 to 340 mu mol/l under applied potential of 200 mV vs. SCE. The sensitivity and detection limit were 90.1 mu A l/mmol cm(2) and 0.3 mu mol/l, respectively. Conclusion: When the sensor was tested in hemolysed erythrocyte samples for GSH determination, a good correlation in results was observed between the sensor and the spectrophotometric method. The sensor showed recovery values between 98% and 102%. (c) 2006 Elsevier B.V All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据