Telemedicine-friendly, portable tonometers: an evaluation for intraocular pressure screening

期刊

CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY
卷 34, 期 7, 页码 666-670

出版社

WILEY
DOI: 10.1111/j.1442-9071.2006.01304.x

关键词

digital; glaucoma; screening; tonometry

向作者/读者索取更多资源

Purpose: To evaluate the intraocular pressure (IOP) readings from two portable, telemedicine-friendly tonometers for suitability in glaucoma screening. Methods: 213 eyes of 107 consenting patients attending an eye clinic were tested with an I-care tonometer and a Pulsair-Easy Eye puff-air tonometer. Gold standard IOP was measured with a Goldmann applanation tonometer (GAT). Effect of central corneal thickness, anterior chamber depth and refractive errors on IOP measurements were also analysed. Results: The mean difference of IOP by GAT and both the portable tonometers was +/- 2.2 mmHg. The analysis indicates minimal difference between IOP readings of both the portable tonometers. The mean difference between two consecutive readings by I-care was 0.01 mmHg. Using 21 mmHg as a threshold for suspected glaucoma, both the portable digital tonometers reported a sensitivity of 38% and specificity of > 95%. In the subjects studied, central corneal thickness had statistically significant influence on IOP measurements while refractive errors and anterior chamber depth had no significant influence on IOP measurements with any tonometry. Conclusions: The IOP readings by both portable tonometers are comparable and were within clinically acceptable range from GAT. These portable tonometers are useful tools for IOP screening.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据