4.7 Article

Determination of selected polychlorinated biphenyls in soil by miniaturised ultrasonic solvent extraction and gas chromatography-mass-selective detection

期刊

ANALYTICA CHIMICA ACTA
卷 577, 期 2, 页码 232-237

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2006.06.050

关键词

micro-scale extraction; ultrasonic solvent extraction; polychlorinated biphenyls; soil; factorial design

向作者/读者索取更多资源

Miniaturised ultrasonic solvent extraction procedure was developed for the determination of selected polychlorinated biphenyls (PCBs) in soil samples by gas chromatography-mass-selective detection by using 2(3) factorial experimental design. Recoveries of PCBs from fortified soil samples are over 90% for three different fortification levels between 40 and 120 mu g kg(-1), and relative standard deviations of the recoveries are below 7%. The limits of detection (LODs) ranged from 0.003 to 0.006 mu g kg(-1). The performance of the proposed method was compared to traditional shake flask extraction method on the spiked real soil sample and extraction methods showed comparable efficiencies. Proposed miniaturised ultrasonic solvent extraction offers several advantages, i.e., reducing sample requirement for measurement of target compound, less solvent consumption and reducing the costs associated with solvent purchase and waste disposal. (c) 2006 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据