4.5 Article

A microfabricated nebulizer for liquid vaporization in chemical analysis

期刊

JOURNAL OF MICROELECTROMECHANICAL SYSTEMS
卷 15, 期 5, 页码 1251-1259

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JMEMS.2006.879671

关键词

fluidics; infrared imaging; mass spectrometry; micromachining; silicon

向作者/读者索取更多资源

A miniaturized nebulizer chip for vaporization of liquid samples for mass spectrometry has been designed, fabricated, and characterized for fluidic and thermal performance. Silicon/glass chip has a liquid sample channel placed centrally between symmetric nebulizer gas channels. The liquid sample is nebulized and vaporised by an integrated platinum heater. The vaporized sample exits through an etched nozzle, and is ionized by an external corona needle. The ions are analysed by a mass spectrometer. The chip has been fabricated in both anisotropically wet etched and DRIE versions in silicon, with an anodically bonded Pyrex glass cover plate. Three different fluidic inlet designs are presented, with both through-wafer and edge insert versions. The shape of the erupting gas jet has been visualized by infrared thermography by using a low-diffusivity imaging screen and high heat capacity helium as a test gas. Dimensions of the jet's thermal footprint on the screen show that the jet is very narrow and confined, and this is confirmed in mass spectrometry results. This confined jet supplies the sample to the ionization region near corona tip, enabling efficient use of very small sample amounts and submicroliter flows.