HILIC mode separation of polar compounds by monolithic silica capillary columns coated with polyacrylamide

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 386, 期 3, 页码 578-585

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-006-0606-z

关键词

monolithic silica; HILIC; polyacrylamide; PA sugars; nucleosides

向作者/读者索取更多资源

HILIC mode columns were prepared by an on-column polymerization of acrylamide on a monolithic silica capillary column modified with N-(3-trimethoxysilylpropyl)methacrylamide as the anchor group. The products showed HILIC mode retention characteristics with three times greater permeability and slightly higher column efficiency compared to a commercially available amide-type HILIC column packed with 5-mu m particles. The selectivity of the monolithic silica-based column was similar to that of the particulate column for each group of solutes towards nucleosides, nucleic bases and carbohydrate derivatives, although a considerable difference was observed in the selectivity for the solute groups. Although the retention of solutes based on the polar functionality was much smaller with the monolithic silica columns, which had a smaller phase ratio, than with the particle-packed column, the former can achieve better separation utilizing the high permeability and higher column efficiencies of a longer column.