Objectives: The aims of this study were to demonstrate the safety and the feasibility of the robotic catheter remote control system (CCS) in endocardial navigation in all cardiac chambers, as well as facilitation of the transseptal puncture. Background: CCS has been developed to facilitate control and precise positioning of catheters within the cardiovascular system. Methods: CCS consists of a remote catheter manipulator, a set up joint, a physician workstation, and a steerable guide catheter (SGC) and sheath. A conventional 4-mm tip catheter was inserted through the SGC to perform mapping of five predefined targets in each cardiac chamber. Seven mongrel dogs were used in this study. Intracardiac echocardiography and three-dimensional (3-D) electroanatomical mapping were integrated with CCS to facilitate catheter manipulation and to guide transseptal puncture. The time to complete the transseptal puncture and the time to complete access to the predefined targets in each cardiac chamber were measured. Gross and microscopic examinations of the accessed and ablation sites were performed to evaluate safety. Results: Transseptal puncture was performed successfully in all animals with a mean time of 7 +/- 3 minutes. Procedure times to access the five targets in the right atrium, right ventricle, left atrium, and left ventricle were 5.6 +/- 1.7, 4.6 +/- 1.5, 13.5 +/- 11.0, 7.0 +/- 2.9 minutes, respectively. There were no intracardiac damages associated with catheter manipulation noted in the excised hearts. Conclusions: Endocardial catheter navigation and mapping using the robotic catheter remote control is safe and feasible. Moreover, the CCS could be used to perform transseptal puncture and left atrial instrumentation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据