4.5 Article

Application of capillary zone electrophoresis with large-volume sample stacking to the sensitive determination of sulfonamides in meat and ground water

期刊

ELECTROPHORESIS
卷 27, 期 20, 页码 4060-4068

出版社

WILEY
DOI: 10.1002/elps.200600166

关键词

capillary zone electrophoresis; large-volume sample stacking; meat; solid phase extraction; sulfonamides; water

向作者/读者索取更多资源

A CZE method with UV-Vis detection has been established and validated for the determination of nine sulfonamides: sulfapyridine, sulfamethazine, sulfamerazine, sulfamether, sulfadiazine, sulfadimethoxine, sulfamethoxazole, sulfachlorpyridazine, and sulfamethizole. Optimum separation was obtained on a 64.5 cm x 75 mu m bubble cell capillary using a buffer containing 45 mM sodium phosphate and 10% methanol at pH 7.3, with temperature and voltage of 27 degrees C and 25 kV, respectively. p-Aminobenzoic acid was used as an internal standard. Taking into account the lack of sensitivity of the UV-Vis detection, the application of an on-line preconcentration methodology, such as large-volume sample stacking with polarity switching has been proposed. This procedure combined with a solvent extraction/SPE method applied for off-line preconcentration and cleanup provides a significant improvement in the LODs, ranging from 2.59 to 22.95 mu g/L for the studied compounds; the quantification of these residues being possible below the levels established by EU legislation in animal food products, such as meat. Satisfactory recoveries were also obtained in the analysis of these compounds in ground water.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据