4.6 Article

Enhancement of an analytical method for the determination of squalene in anthrax vaccine adsorbed formulations

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.jpba.2006.04.009

关键词

squalene; anthrax vaccines; Gulf War syndrome; reverse-phase liquid chromatography; method validation; detection limits

向作者/读者索取更多资源

Specific lots of anthrax vaccine adsorbed administered to members of the U.S. Armed Forces have been alleged to contain squalene, a chemical purported to be associated with illnesses of Gulf War veterans. A method of enhanced sensitivity for determining squalene in anthrax vaccine adsorbed using high-performance liquid chromatography with photodiode array detection has been developed, validated, and applied to 44 bottles of 38 lots of anthrax vaccine. In 43 bottles of 37 lots, no squalene was detected within a detection limit of 1 ng/0.5 ml dose (2 parts-per-billion). One lot, FAV008, was found to contain trace amounts of squalene at 7, 9, and 1 mu g l(-1), levels considerably below nomial human plasma levels (290 mu g l(-1)). The overall results of this investigation provide direct evidence for the absence of squalene in nearly all of anthrax vaccine preparations tested. (c) 2006 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据