4.4 Article

Assessing the capabilities of a 4kx4k CCD camera for electron cryo-microscopy at 300kV

期刊

JOURNAL OF STRUCTURAL BIOLOGY
卷 156, 期 3, 页码 556-563

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jsb.2006.08.019

关键词

cryo-EM; CCD camera; 300 kV; electron cryo-microscopy

资金

  1. NCRR NIH HHS [P41RR02250] Funding Source: Medline

向作者/读者索取更多资源

CCD cameras have numerous advantages over photographic film for detecting electrons; however the point spread function of these cameras has not been sufficient for single particle data collection to subnanometer resolution with 300 kV microscopes. We have adopted spectral signal to noise ratio (SNR) as a parameter for assessing detector quality for single particle imaging. The robustness of this parameter is confirmed under a variety of experimental conditions. Using this parameter, we demonstrate that the SNR of images of either amorphous carbon film or ice embedded virus particles collected on a new commercially available 4kx4k CCD camera are slightly better than photographic film at low spatial frequency (< 1/5 Nyquist frequency), and as good as photographic film out to half of the Nyquist frequency. In addition it is slightly easier to visualize ice embedded particles on this CCD camera than on photographic film. Based on this analysis it is realistic to collect images containing subnanometer resolution data (6-9 angstrom) using this CCD camera at an effective magnification of similar to 112000 x on a 300 kV electron microscope. (c) 2006 Elsevier Inc. All rights reserved.