4.4 Article

Survival of NobelDirect implants:: An analysis of 550 consecutively placed implants at 18 different clinical centers

期刊

出版社

WILEY
DOI: 10.1111/j.1708-8208.2007.00054.x

关键词

dental implants; failures; one-piece implant; retrospective clinical study

向作者/读者索取更多资源

Background: A recently introduced implant, the NobelDirect (Nobel Biocare AB, Goteborg, Sweden), has previously been documented with substantial bone resorption in a large number of operated cases. Purpose: The aim of this study was to evaluate the failure rate of NobelDirect implants in a retrospective multicenter survey. Materials and Methods: A total of 550 NobelDirect implants consecutively placed in over 269 patients at 18 centers were evaluated with regard to failure rate after an average follow-up of about 1 year. Results: The overall failure percentage was 10.9% (59 failures). The 58 implants not loaded directly showed only one failure (1.7%) versus 58 failures (11.8%) of those implants that were loaded directly. A chemical x-ray photoelectron spectroscopy analysis of an implant from the original batch showed up to 3.5% silicon at parts of the implants. A retrieval analysis of one implant removed at 2 years after placement demonstrated bone resorption down to the level of the fifth thread. Conclusions: It is concluded that the NobelDirect implant, if placed with a punch procedure, ground down in situ, and loaded directly, shows an unusually high failure rate at 1 year.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据