4.6 Article

Usefulness of mechanical transvenous dilation and location of areas of adherence in patients undergoing coronary sinus lead extraction

期刊

EUROPACE
卷 9, 期 1, 页码 69-73

出版社

OXFORD UNIV PRESS
DOI: 10.1093/europace/eul130

关键词

resynchronization therapy; lead extraction

向作者/读者索取更多资源

Aims Few data have been currently reported on the outcome of coronary sinus (CS) lead removal, particularly using mechanical dilation (MD). We aimed to evaluate feasibility, safety, and effectiveness of CS lead extraction, focusing on MD usefulness, in the event that lead traction (LT) was ineffective. Methods and results We studied 37 consecutive patients (30 mates, mean age 68.1, range 52-80), who underwent left ventricle (LV) pacing lead removal; the indication for extraction was local infection in 16 patients (43.3%), sepsis in 11 patients (29.7%), and lead malfunction in 10 patients (27%). The procedure was first attempted by LT, followed, if unsuccessful, by MD using polypropylene sheaths. All CS leads (time from implant 19.5 +/- 16.5, range 2-84 months) were successfully removed; LT was effective (LT group) in 27 patients (73%) and ineffective in 10 patients (27%), for whom MD was necessary (MD group). There were no major complications. The areas of adherence were in the CS in only one patient. No differences were noted in the data analysed between LT and MD groups; in particular, time from implant was similar in the two groups (MD vs. LT group: 17 +/- 8.9 vs. 20.4 +/- 18.6 months; P = ns). Conclusion Our study suggests that CS leads, after medium-term implantation, can be effectively and safety removed using MD with polypropylene sheaths, in the case of unsuccessful LT. No preoperative elements predictive of LT failure could be identified. Areas of adherence were rarely located in the CS or its tributaries.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据