4.7 Article

Transjugular liver biopsy: prospective evaluation of the angle formed between the hepatic veins and the vena cava main axis and modification of a semi-automated biopsy device in cases of an unfavorable angle

期刊

EUROPEAN RADIOLOGY
卷 17, 期 1, 页码 169-173

出版社

SPRINGER
DOI: 10.1007/s00330-006-0256-8

关键词

liver biopsy; liver, interventional procedures; transjugular liver biopsy; biopsy technology

向作者/读者索取更多资源

In cases of transjugular liver biopsies, the venous angle formed between the chosen hepatic vein and the vena cava main axis in a frontal plane can be large, leading to technical difficulties. In a prospective study including 139 consecutive patients who underwent transjugular liver biopsy using the Quick-Core biopsy set, the mean venous angle was equal to 49.6 degrees. For 21.1% of the patients, two attempts at hepatic venous catheterization failed because the venous angle was too large, with a mean of 69.7 degrees. In all of these patients, manual reshaping of the distal curvature of the stiffening metallic cannula, by forming a new mean angle equal to 48 degrees, allowed successful completion of the procedure in less than 10 min.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据