Determination of morphine and codeine in urine using poly(dimethylsiloxane) microchip electrophoresis with electrochemical detection

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.jpba.2006.06.003

关键词

poly(dimethylsiloxane) (PDMS) microchip; electrochemical (EC) detection; morphine; codeine; beta-cyclodextrin (P-CD); acetonitrile

向作者/读者索取更多资源

In this paper, a poly(dimethylsiloxane) (PDMS) microchip with electrochemical (EC) detection was developed for rapid separation and detection of morphine and codeine. It was found that morphine and codeine were well separated within 140 s in phosphate buffer solution (PBS) (pH 6.6, 40 mM)-beta-cyclodextrin (beta-CD) (20 mM)-acetonitrile (30%, v/v). The detection limit was 0.2 mu M for morphine and 1 mu M for codeine. The protocol was successfully applied to monitoring the amount of morphine and codeine in human urine. Compared with the conventional methods, the presented method had many advantages such as lower instrument cost, less reagent consumption and shorter analysis time. (c) 2006 Elsevier B.V. All rights reserved.