4.4 Article

A new europium(III) PVC membrane potentiometric sensor based on 4-(2-hydroxybenzylideneamino)-6-methyl-3-thioxo-3,4-dihydro-1,2,4-triazin-5(2H)-one

期刊

BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN
卷 80, 期 1, 页码 172-177

出版社

CHEMICAL SOC JAPAN
DOI: 10.1246/bcsj.80.172

关键词

-

向作者/读者索取更多资源

A new europium ion-selective electrode based on 4-(2-hydroxybenzylideneamino)-6-methyl-3-thioxo-3,4-dihydro-1,2,4-triazin-5(2H)-one (HMTDT) as membrane carrier was prepared. The electrode has a wide concentration range between 1.0 x 10(-1) and 1.0 x 10(-6) mol dm(-3) with a near-Nernstian slope of 19.7 +/- 0.4 mV per decade and a detection limit of 7.8 x 10(-7) mol dm(-3). The potentiometric response is independent of the pH of the solution in the pH range 3.5-7.8. It has a relatively fast response time and can be used for at least 2 months without any considerable divergence in potentials. The proposed sensor had very good selectivities for a wide variety of metal ions. The practical utility of the electrodes has been demonstrated by their use as indicator electrodes in the potentiometric titration of Eu3+ ions with EDTA and for the determination of Eu3+ in different solutions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据