4.3 Article

In vitro frictional forces generated by three different ligation methods

期刊

ANGLE ORTHODONTIST
卷 78, 期 5, 页码 917-921

出版社

E H ANGLE EDUCATION RESEARCH FOUNDATION, INC
DOI: 10.2319/090607-420.1

关键词

friction; self-ligating bracket; low-friction biomechanics; elastomeric ligatures

向作者/读者索取更多资源

Objective: To test the hypothesis that there is no difference between the frictional forces produced by a passive self-ligating bracket (SLB) in vitro and a conventional bracket (CB) used with two types of elastomeric ligatures. Materials and Method: The brackets, wires and ligation methods used in vitro were a passive SLB and a CB used with two types of elastomeric ligatures (conventional elastomeric ligature [CEL] and unconventional elastomeric ligatures [UEL]). The bracket ligation systems were tested with two types of wires (0.014 '' super elastic nickel titanium wire and 0.019 '' x 0.025 '' stainless steel wire). Resistance to sliding of the bracket/wire/ligature systems was measured with an experimental model mounted on the crosshead of an Instron testing machine with a 10 N load cell. Each sample was tested 10 consecutive times under a dry state. Results: Frictional forces close to 0 g were recorded in all tests with SLB and in all tests with UEL on CB with both wire types. Resistance to sliding increased significantly (87-177 g) (P < .05) when CEL on CB was used with both wires. Conclusion: UELs may represent a valid alternative to passive SLBs for low-friction biomechanics.