期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 846, 期 1-2, 页码 105-111出版社
ELSEVIER
DOI: 10.1016/j.jchromb.2006.08.026
关键词
pseudoephedrine; cetirizine; HPLC-MS/MS; human plasma; pharmacokinetics
A sensitive, simple and highly selective liquid chromatography/tandem mass spectrometry (LC/MS/MS) method was developed and evaluated to determine simultaneously the concentrations of pseudoephedrine and cetirizine in human plasma. The chief benefit of the present method is the minimal sample preparation, as the procedure is only one-step protein precipitation. Two drugs were separated on a C-8 column and analyzed by LC/MS/MS using positive electrospray ionisation (ESI). The method had a chromatographic run time of 12.0 min and a linear calibration curve over the concentration range of 1.0-800 ng/ml for pseudoephedrine and 1.0-400 ng/ml for cetirizine, respectively. The lower limit of quantification of the two drugs was 1.0 ng/ml, respectively. The intra- and inter-batch precisions were less than 9.7%. The method described herein has been first used to reveal the pharmacokinetic characters in healthy Chinese volunteers treated with oral administration of different dosages of cetirizine dihydrochloride and controlled-released pseudoephedrine hydrochloride compound tablet, and approached the influence of a standard meal on the extent and rate of absorption of the combination tablet. (c) 2006 Elsevier B.V. All rights reserved.