期刊
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 54, 期 1, 页码 3-10出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2006.890013
关键词
image analysis and processing; NMIS; PET instrumentation; reconstruction algorithms
A growing interest in the development of dual modality PET/CT scanners prompts the comparative study of numerous scintillators to select the best one, which could be used simultaneously in PET detectors working in the pulsing mode and in the CT detectors working in the current mode. In the comparative measurements, done in the same experimental conditions, various samples of BGO, GSO, GSO:Ce, Zr, LGSO, LSO, LYSO, MLS, LaC13, LaBr3 and CWO scintillators were tested. The measurements covered a determination of the light output, energy resolution, non-proportionality of the light yield, decay times of the light pulses and for the selected crystals their time resolution for 511 keV annihilation quanta. Moreover, a comparative study of afterglow, induced by 60 keV gamma-rays from a strong Am-241 source (13.9 GBq), was done in the second range of time. The LSO-like crystals are the best in the PET scanners application. However, they do not fit to the CT requirements, due to a ig afterglow. The studies conclude that besides of the well known BGO, only GSO:Ce and most likely LaBr3 might be considered for the simultaneous PET/CT detector.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。