期刊
MEDICAL PHYSICS
卷 34, 期 2, 页码 689-702出版社
WILEY
DOI: 10.1118/1.2409480
关键词
-
资金
- NCI NIH HHS [R21 CA098691, R01 CA119056] Funding Source: Medline
We studied the performance of a dual-panel positron emission tomography (PET) camera dedicated to breast cancer imaging using Monte Carlo simulation. The PET camera under development has two 10 x 15 cm(2) plates that are constructed from arrays of 1 x 1 x 3 mm(3) LSO crystals coupled to novel ultra-thin (< 200 mu m) silicon position-sensitive avalanche photodiodes (PSAPD). In this design the photodetectors are configured edge-on with respect to incoming photons which encounter a minimum of 2 cm thick of LSO with directly measured photon interaction depth. Simulations predict that this camera will have 10-15% photon sensitivity, for an 8-4 cm panel separation. Detector measurements show similar to 1 mm(3) intrinsic spatial resolution, < 12% energy resolution, and similar to 2 ns coincidence time resolution. By performing simulated dual-panel PET studies using a phantom comprising active breast, heart, and torso tissue, count performance was studied as a function of coincident time and energy windows. We also studied visualization of hot spheres of 2.5-4.0 mm diameter and various locations within the simulated breast tissue for 1 x 1 x 3 mm(3), 2 x 2 x 10 mm(3), 3 x 3 x 30 mm(3), and 4 x 4 x 20 mm(3) LSO crystal resolutions and different panel separations. Images were reconstructed by focal plane tomography with attenuation and normalization corrections applied. Simulation results indicate that with an activity concentration ratio of tumor: breast: heart: torso of 10:1:10:1 and 30 s of acquisition time, only the dual-plate PET camera comprising 1 x 1 x 3 mm(3) crystals could resolve 2.5 mm diameter spheres with an average peak-to-valley ratio of 1.3. (c) 2007 American Association of Physicists in Medicine.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据