High antinoise photoacoustic tomography based on a modified filtered backprojection algorithm with combination wavelet

期刊

MEDICAL PHYSICS
卷 34, 期 2, 页码 556-563

出版社

WILEY
DOI: 10.1118/1.2426406

关键词

photoacoustic tomography; combination wavelet; antinoise capacity; noise; filtered backprojection

向作者/读者索取更多资源

How to extract the weak photoacoustic signals from the collected signals with high noise is the key to photoacoustic signal processing. We have developed a modified filtered backprojection algorithm based on combination wavelet for high antinoise photoacoustic tomography. A Q-switched-Nd: yttrium-aluminum-garnet laser operating at 532 nm is used as light source. The laser has a pulse width of 7 ns and a repetition frequency of 20 Hz. A needle polyvinylidene fluoride hydrophone with diameter of 1 mm is used to capture photoacoustic signals. The modified algorithm is successfully applied to imaging vascular network of a chick embryo chorioallantoic membrane in situ and brain structure of a rat brain in vivo, respectively. In the reconstructed images, almost all of the capillary vessels and the vascular ramifications of the chick embryo chorioallantoic membrane are accurately resolved, and the detailed brain structures of the rat brain organization are clearly identified with the skull and scalp intact. The experimental results demonstrate that the modified algorithm has much higher antinoise capacity, and can greatly improve the reconstruction image quality. The spatial resolution of the reconstructed images can reach 204 mu m. The modified filtered backprojection algorithm based on the combination wavelet has the potential in the practical high-noise signal processing for deeply penetrating photoacoustic tomography. (c) 2007 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据