4.7 Article

Flow-injection chemiluminescence determination of formaldehyde in water

期刊

TALANTA
卷 71, 期 2, 页码 900-905

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2006.05.078

关键词

formaldehyde; chemiluminescence; flow-injection analysis; simplex

向作者/读者索取更多资源

A modification of the Trautz-Schorigin reaction into a flow-injection analysis configuration is described. Different approaches were used at the optimization of chemi luminescence determination of formaldehyde in water based on the reaction of formaldehyde, gallic acid and hydrogen peroxide in an alkaline solution. Detection system with a 218 mu l chemiluminescence cell was optimized by both a one-variable-at-a-time method, and a modified simplex method. A calibration graph is linear in the concentration range 4 x 10(-8) to 1 x 10(-5) M HCHO. The detection limit of formaldehyde for a signal-to-noise ratio of 3 is 4 x 10-8 M. The relative standard deviations for 15 repeated measurements of 1 x 10(-6) and 5 x 10(-6) mol l(-1) HCHO are 4.32 and 3.33%, respectively. The analysis time is 1.5 min. The method was applied to the determination of formaldehyde in urban rainwater. A comparison of results found by proposed method with those obtained by fluorimetric reference method provided a good agreement. (c) 2006 Elsevier B.V. All rights reserved.